1.1 This Part of CEN/TS 14507 applies to systems for the supply of nitric oxide/nitrogen mixtures to a
delivery and monitoring system complying with CEN/TS 14507-1 for treatment of one patient at a time, in a
healthcare facility.
1.2 This Part of CEN/TS 14507 applies to supply systems with cylinders at a filling pressure up to 20 000
kPa and concentrations up to 1000 μl/l of nitric oxide in nitrogen.
1.3 R This Part of CEN/TS 14507 does not apply to:
· pipeline systems for delivery of nitric oxide/nitrogen mixtures to more than one patient at a time;
· supply systems for nitric oxide/nitrogen mixtures intended for use in home care, emergency and transport.